TSS Turbidity Analyzer

TSS turbidity analyzer can be used for position detection of the sewage treatment plant and other related plant water inlet and water outlet turbidity.
● Sewage treatment plant measurement of inlet and outlet concentration of suspended solids
● Turbidity measurement of backwash water in waterworks and measurement of turbidity of raw water and sedimentation tank effluent
● Water quality of industrial production process/circulating cooling water

The absorption, reflection and scattering of the infrared light emitted by the light source during the transmission process. In a certain range of scattering light intensity and turbidity / suspended matter concentration is proportional to TSS, it is the basic principle of 135 degrees and 90 degrees respectively in the direction of two sets of light receivers based on, through the analysis of the two groups receiving light intensity so that the turbidity value or suspension concentration value.

● It can not only detect the turbidity, but also the concentration of suspended substance (or sludge);
● Double beam infrared detection technology;
● Digital sensor, with strong anti-interference ability, sensor cable length up to several hundred meters;
● Sensor plug and play, automatic identification;
● Analyzer can provide a way to isolate the RS232 digital interface and isolated RS485 digital interface, support standard Modbus ASCII and Modbus RTU protocol;
● Support two languages: Chinese and English;
● Relay output programmable: any relay can be set to high alarm, low alarm, timer mode;
● With a history of automatic data storage, storage, query 10 years of historical data

Measurement performance
● Measuring range: 0~4000NTU,0~1000mg/L
● Resolution: 0.01NTU,0.01mg/L
● Accuracy: Reading±2%,10NTU error is not greater than 1NTU
● Calibration: Factory calibration, on-site calibration
● Display: LCD liquid crystal display date, time, measurement, historical trend line, Chinese and English menu options
● Data storage: automatic storage function, power failure data protection, storage, query 10 years of historical data
● Site settings: completed by the transmitter button
Output
● Analog output: analog output current isolation 0/4~20mA, the maximum load of 750, the fault state output current optional
● Relay output: 3 programmable relays can be set to alarm function and timing function
● Relay capacity: 2A, 250VAC
● Digital interface: MODBUS RS232/RS485, Profibus DP optional
Power supply
● AC power supply: 85~265VAC,50Hz/60Hz
● DC power supply: 24VDC±10%
Physical property
● Transmitter size: 270×205×100mm
● Sensor size: Φ60×155mm
● Material: transmitter: ABS+PC sensor: 316L
● Connection: 1 inch taper external thread
Environmental performance
● Protection level: transmitter: IP65 sensor: IP68
● Working temperature: transducer: -20 ~60, Sensor: 0 ~60
● Relative humidity: 0~95%, no condensation
● Working pressure: maximum 6Bar